Temperature Controllers
|
|
TC5 Temperature Controller |
TC8 Temperature Controller |
Click on any drawing on this page to see enlarged version.
(Must have Adobe Reader to view these PDF drawings.)

TC5 Benchtop Temperature Controller with Built-in Power Supply
The Model TC5 Temperature Controller allows safe, convenient operation of the thermoelectrically cooled detectors. The TC5 Controller is easy to use. The built-in power supply accepts AC line voltage. Detector temperature is set on the controller front panel, and a digital meter indicates current drawn by the cooler. The current-overload feature protects the cooler from damage due to overheating; cooler power is cut off if the cooler attempts to draw excess current (e.g. due to improper heat sink, improper cooler input polarity, or defective thermistor.) The controller's connector for thermistor input and cooler output mates to the HS1 Heat Sink Assembly.
Key Features
- Proportional/Integral Temperature Control
- 30 Watt Capacity
- Temperature Select Front Panel
- Easy Connections to HS1 Heat Sink/Cable Assembly
- Built-in Power Supply
- Built-in Protection against TE cooler damage
- ±15V Output for Amplifier Circuit

Click for more TC5 information.
TC8 Temperature Controller
The Model TC8 is a self-contained Temperature Controller for use with the thermoelectric coolers. Using a single 5 Volt power supply, the TC8 operates in conjunction with a thermistor, located in the detector assembly, to precisely measure and regulate the temperature of the cooler detector. The TC8 is also available with detector heat sink and optional hybrid amplifier in the CM21 and the CMAMP assemblies.
Click for more TC8 information.
|
Model |
TC5 |
TC8 |
Supply
Voltage |
120V
AC Line Voltage
(220V optional) |
+5.0
± 0.25 VDC @ 2.0A |
Maximum
Current to TE Cooler |
2A @
15V |
1.5A @
5V |
Detector
Temperature |
+10°C
to -80°C |
Ambient
to -80°C |
Temperature
Stability |
±0.5°C |
±
0.5°C |
Ambient
Temperature |
+10°C
to +40°C |
+10°C
to +40°C |
Stabilized
Temperature Indication |
Front
Panel Multimeter |
LED |
Recommended
for Detectors: |
All
Judson TE cooled detectors |
All
Judson TE cooled detectors |
Heat Sink Assemblies
|
|
CMAMP Assembly |
HS1 Assembly |
A thermoelectrically cooled detector requires a heat sink to dissipate the heat generated by the cooler, an amplifier to amplify the detector signal to a usable level and a temperature controller to hold the detector at a constant temperature. Judson TE cooler accessories are designed to give solutions for these problems to our customers.
HS1 Assembly (Heat Sink Assembly for TO-37, TO-66 and TO-3 Packages)
The HS1 series is designed to provide heat sinking and cabling to simplify integrating a TE cooled detector into an existing system, and is designed to mate with either the TC-5 or TC-8 temperature controllers.
HSAMP Assembly (Heat Sink and Amplifier Assy for TO-37, TO-66 and TO-3 Pkgs)
The HSAMP series is designed to provide both heat sinking and signal amplification for our TE cooled photovoltaic product line. This allows Judson to help a customer to choose the best amplifier for a particular detector and system. The HSAMP includes a hybrid PA-5, PA-6 or PA-7 preamplifier and is designed for easy connection to the customer's optical system and the TC-8 Temperature Controller. 
CM21 Assembly (Heat Sink and TC8 Temperature Controller for TO-37, TO-66 and TO-3 Pkgs)
The CM21 assembly is designed to provide heatsinking and temperature control for Judson's TE cooled photoconductive and photovoltaic detector product line. It is designed to simplify integrating a TE cooled detector into an exisiting system with a customer supplied amplifier. The CM21 assembly is ideal for customers who have dsigned their own amplifier for their system.
CMAMP Assembly (Heat Sink Amplifier and TC8 Temperature Controller for TO-37, TO-66 and TO-3 Pkgs)
The CMAMP series is designed for customers that would like a fully integrated detector module. It includes heatsinking, detector signal amplification and temperature control. The CMAMP assembly requires an external power supply that provides both +5V @ 2.0 amperes and ±15 volts at 100mA. The CMAMP assembly is an ideal platform for evaluating TE cooled detectors for the first time. Custom mechanical assemblies including temperature control, detector amplification and heat sinking are available on an OEM basis.

Click here to view table

Click here to view table
Click for more information on cooler accessories.
|